Site Map | 中文版 | Русский |   
 
不支持flash
 
   
 
 Home - Products & Systems - Three Phase Energy Meter Test Bench HS-6303
Three Phase Energy Meter Test Bench HS-6303
Product Group: Energy meter equipments
Product Description

Features

1)        The bench accords with the standard of “IEC 60736:1982 testing equipment for electrical energy meters”, “IEC 62053”

2)        The bench can be used to test both single/three phase active energy meter and reactive energy meter (up to 0.2 class), it is also suitable for four quadrant energy meter testing.

3)        The bench can do warm up test, starting test, no load test(creep test), error test(accuracy test), dial test, tests of the influence quantities(reverse phase, voltage unbalance, harmonic wave and so on) and test of repeatability of the measurements.

4)        The bench is using the PWM to amplify voltage/current output, the high efficiency at >85%, and lower the heat. The PWM can check the malfunction checking, preset, alarm, and protection

5)        The signal source is using the newest digital controlled sine wave technology for frequency, amplitude and phase adjust. The signal source can repeatedly add 2-21 harmonics on the first sine wave, or cause sub-harmonic wave and odd sub-harmonic wave.

6)        The bench is using the HY 5303C-22 reference standard energy meter with comparison technology to offer the accuracy of energy meters.

7)        The bench is using the error processor system to calculate the error of each energy meter by the sub-error system, which can be connected by the internal RS485.

8)        The same specification but different constant energy meter can be tested at one time.

9)        The test bench can be installed scanning head, it can offer 3 ranges adjustment. It is also can be accepted the signal for both round-style energy meter and digital LED energy meter.

10)    The test bench has the auto-search black marker function which can precisely locate the black marker during the no load test and starting test.

11)    The bench can choose to operate by keyboard or PC which is very convenient. The working status of bench can be seen very clear and easy operation.

12)    The bench is using the testing software complied with WINDOWS98/2000/XP which can auto-testing different energy meters.

13)    The software can realize the data from energy meter for statistics, search and management.

14)    The energy meter hanging rack is made by Aluminum compound. The whole device configuration is a reasonable design and nice outfit

Technical Specification

1)        The test bench accuracy class: 0.05/0.1

2)        Reference standard energy meter:

          HY 5303C-22 Three Phase Reference Standard Meter

          Accuracy class: 0.05 /0.1

3)        Voltage amplifier:

Output voltage range: 0-300V(phase-neutral);

Output maximum power: 800VA

Resolution: 0.01%;

Stability: better than 0.05%/3 min(time base 5S);

Distortion factor: < 0.5% at linear load;

Frequency range: 45Hz~65Hz;

Power efficiency: more than 85%;

Ambient temperature:10℃~50℃

Electronic overload and short-circuit protection.

4)        Current amplifier:

Output current: range: 1mA~120A (option for 1mA~200A);

Output maximum power: 1500VA

Resolution: 0.01%;

Stability: better than 0.05%/3 min(time base 5S);

Distortion factor: < 0.5% at linear load;

Frequency range: 45Hz~65Hz;

Power efficiency: greater than 85%;

Ambient temperature:10℃~50℃

Electronic overload protection and security for open current circuit.

5)        Phase Angle: 0°~360° , resolution of phase angle setting: 0.1°;

6)        Output frequency range: 45Hz~65Hz; resolution: 0.01Hz

7)        Harmonic wave output: 2-21 sub-harmonic wave     contents≤40%

                     3, 5, 7 sub-harmonic wave coupling, total content≤40%

                         Phase angle control (odd sub-harmonic wave)

                         Wave group control (sub-harmonic wave)

8)        Output stability of power: <0.05% (180S)

9)        Numbers of calibrated energy meter: 16 (6/8/12/16/20/24/32 on requested)

10)    Input power: 3×220V/380V (Y) 15%, 50/60Hz; Max. power consumption: 6000VA

11)    Dimension:

Type of bench

Meters

Number of rack

Number of row

Size of rack

Integration

6

8

12

16

24

32

1

1

1

1

2

2

Single

Single

Double

Double

Double

Double

2000×800×1600 mm

2200×800×1600 mm

2000×800×1920 mm

2200×800×1920 mm

2000×800×1920 mm

2200×800×1920 mm

Fission

6

8

12

16

24

32

1

1

1

1

2

2

Single

Single

Double

Double

Double

Double

2000×700×1600 mm

2200×700×1600 mm

2000×700×1920 mm

2200×700×1920 mm

2000×700×1920 mm

2200×700×1920 mm

Size of the Console: 600×800×1920 mm

Products
 
AMR system
Prepayment system
ZigBee
Electronic meter
Mechanical meter
Energy meter equipments
Gas Meter
Water Meter
 
Contact us

No.181 Wuchang Avenue,
Yuhang District, Hangzhou 310023,
Zhejiang, China

Tel: 86-571-89300666/89300575
Fax: 86-571-89300777

 
 
 
 
 